    Standard Test Areas in Textile and Membrane Testing

    By Scientific Converters · Updated · Factors from NIST SP 811 and the BIPM SI Brochure

    Short answer: ISO 9237 uses a 20 cm² test area (with 5, 50 and 100 cm² alternatives) and ASTM D737 uses 38.3 cm² (about 5.94 in²); permeability is the measured flow divided by that area.

    The common test areas

    Air permeability and filter-media testers clamp the sample over a circular opening, the test head. The area of that opening turns the measured flow into a permeability. These areas cover nearly all commercial testers:

    Standard test-head areas with exact conversions
    Areain²ft²Diameter (mm)Where used
    5 cm² 0.775 in² 0.005382 ft² 25.23 ISO 9237 (alternative test area)
    10 cm² 1.55 in² 0.01076 ft² 35.68 Instrument option on many permeability testers
    20 cm² 3.1 in² 0.02153 ft² 50.46 ISO 9237 / EN ISO 9237 (standard test area)
    25 cm² 3.875 in² 0.02691 ft² 56.42 Instrument option; used for nonwovens and small samples
    38.3 cm² 5.937 in² 0.04123 ft² 69.83 ASTM D737 (standard test area, ≈ 5.94 in²)
    50 cm² 7.75 in² 0.05382 ft² 79.79 ISO 9237 (alternative test area)
    100 cm² 15.5 in² 0.1076 ft² 112.8 ISO 9237 (alternative); filter media holders (e.g. 32 L/min → 5.33 cm/s)

    Area and diameter of a round test head

    For a circular opening, A = π d² / 4, so d = √(4A/π). A 20 cm² head is 50.46 mm across and a 38.3 cm² head 69.83 mm. Measure the inner diameter of the clamping ring when checking a tester; a 1 % error in diameter is a 2 % error in area and therefore in permeability.

    Why the area changes the flow you read

    At the same permeability, flow scales with area. A fabric of 100 mm/s passes 12 L/min through 20 cm² but about 22.98 L/min through 38.3 cm². Choose a head that keeps the flow in the accurate range of the tester's flow meter.

    Choosing an area

    • Use the area the specification names; results on different areas should agree, but specifications are written for one.
    • Use larger areas (50–100 cm²) for open or variable materials such as nonwovens, so a single thin spot does not dominate.
    • Use small areas (5–10 cm²) for small samples or very open materials that would exceed the flow range.

    Frequently asked questions

    What test area does ISO 9237 use?

    ISO 9237 specifies 20 cm² as the standard test area, with 5, 50 and 100 cm² allowed as alternatives.

    What test area does ASTM D737 use?

    ASTM D737 uses 38.3 cm², which is 5.937 in².

    Does the test area change the permeability result?

    Ideally no: permeability is flow per area, so a larger head simply passes more air. In practice small areas are more sensitive to edge leakage and local variation in the material.